| Model Number |
EGIATRS45AXT |
| Medical Device Problem Codes |
Break (1069); Component or Accessory Incompatibility (2897); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intraoperatively, the reload was unable to be attached to handle.The handle has been used three times prior.Assistance was given to the scrub nurse to try and attach the reload, but after several tries, a new reload was opened and used without issue.A piece of black plastic was later found dislodged from the connection end of the reload.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual examination of the staple cartridge noted that the reload had a full complement of staples.The lock ring was observed to be broken.Functionally the reload was loaded into a pmv representative instrument and was applied to test media with proper staple placement and media transection.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the damaged reload lock ring may occur due improper orientation of the lock ring or over flexure of the reload while attached to the instrument.The root cause of the observed damage was due to the product not being used as intended which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|