ETHICON ENDO-SURGERY, LLC. PROXIMATE*HCS HEMORR CIR STAPL; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number PPH01
Health Effect - Clinical Codes Fever (1858); Pain (1994); Septic Shock (2068); No Code Available (3191)
Date of Event 01/01/2004
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Publication year of 2004.Batch # unk.This report is related to a journal article; therefore, no product will be returned for analysis and the manufacturing records cannot be reviewed as the lot/batch number has not been provided.(b)(4).
 
Event or Problem Description
It was reported via literature entitled: pelvic sepsis after stapled hemorrhoidectomy authors: patrick pessaux, md; emilie lermite, md; jean-jacques tuech, md; olivier brehant, md; nicolas regenet, md; jean-pierre arnaud, md citation: j am coll surg.2004; 199(5): 824-5.Doi: 10.1016/j.Jamcollsurg.2004.04.026.A previously healthy (b)(6) year-old man underwent stapled hemorrhoidectomy done with pph 01 33-mm ethicon hemorrhoidal circular stapler (ethicon).Four days later, the patient was presented back at the emergency department complaining of rectal discomfort and perineal pain.The patient had septic shock and fever of 39.2 °c.Perineal examination showed extensive subcutaneous necrosis of the scrotum and inguinal area with crepitus on palpation.Ct showed a leakage from the staple line with extensive gas developing the rectum, bladder, and scrotum.The patient underwent emergency surgery.The rectum was inflated with water and povidone-iodine.Debridement and an end-colostomy were performed.Repeat necrosectomy and hyperbaric oxygen therapy were done during the stay.Culture of the necrotic tissue showed a polymicrobial infection (escherichia coli, bacteroides fragilis, and enterococci).The wound healed completely by 3 months.The patient presented with anal stenosis that required two dilations.It was reported that the perforation may have been a delayed ischemic phenomenon, related to crushing of the rectal wall by closure of the gun before final stapling.Usually the infection is polymicrobial, frequently with gas-forming organisms.It is possible that the firing of the stapler enabled gas-producing organisms in the rectal lumen to enter the pararectal space.Care must be taken to avoid incorporating more proximal prolapsing rectal mucosa into the jaws of the closing gun.Patients with anoreceptive sexual practices require appropriate advice.It is debatable whether anoreceptive sexual practice is a contraindication to stapled hemorrhoidectomy.Individuals should be warned that the presence of a circumferential staple line at the anorectal junction may make anal intercourse intolerable for both parties.The ¿go ahead¿ for anal sex should be given only after rectal digitation by the surgeon has confirmed that the staple line is correctly healed.Although this complication is exceptional, it must be included in consent information given to patients who are treated for hemorrhoid symptoms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key8842960
Report Number3005075853-2019-20897
Device Sequence Number14090479
Product Code GDW
UDI-Device Identifier10705036012993
UDI-Public10705036012993
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K030925
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source literature
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberPPH01
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/11/2019
Initial Report FDA Received Date07/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-