| Catalog Number |
364816 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/27/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd received samples from the customer facility for investigation.The samples were tested/evaluated and the customer's indicated failure mode for stopper pullout with the incident lot was not observed as all product specifications were met.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Investigation conclusion: based on evaluation of the customer samples, the customer¿s indicated failure mode for stopper pullout with the incident lot was not observed as all samples met the required specifications.Root cause description: based on the investigation, a root cause could not be determined.The product was found to be in conformance and meet release specifications.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that bd vacutainer® acd solution b blood collection tubes stopper pulls out.This was discovered during use.The following information was provided by the initial reporter: material no.364816, batch no.9126927.It was reported that the 2nd stopper pullout force did not meet our specification.During r&d testing in (b)(4), we identified an instance where the 2nd stopper pullout force did not meet our specification.Test product: catalog # 364816, batch 9126927.Test date: (b)(6) 2019.Data review date: july 9, 2019.
|
| |
|
Search Alerts/Recalls
|