| Model Number |
RONYX40018UX |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Myocardial Infarction (1969)
|
| Date of Event |
12/07/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
During the index procedure two resolute onyx des were implanted in the rca.The patient suffered a target vessel myocardial infarction due to in- stent restenosis and acute thrombosis on the same day and revascularisation was performed by implanting two resolute onyx des in the rca.Cec adjudicated the mi a non q wave target vessel mi 3rd udmi peri-pci and the revasc as clinically driven tlr.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: after the index procedure the patient had st elevation and chest pain which resolved at presentation to the cath lab.Revascularisation was performed by implanting two resolute onyx des in the rca on the same day.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|