|
Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime or a33 test, excessive no delivery test and delivery accuracy test.The stop (idle) current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and a21 error test.The no delivery alarm functions properly during the basic occlusion test and occlusion test.No unexpected no delivery alarm noted.The test battery was removed and reinserted with no unexpected insulin pump reset or a21 alarm noted.No lost settings or time clock anomaly noted during testing.Device communicated properly with the test blood glucose meter.The test value of 95 mg/dl was sent by the meter and received by the device.No insulin pump or meter communication anomaly noted.Device uploaded properly using carelink.No anomaly noted when uploading the insulin pump using carelink.Carelink properly listed the test data.Device was monitored for 1 day.No unexpected alarms noted during testing.No cracked keypad overlay noted.No damage noted on the lcd glass.Device had cracked reservoir tube near the esc and act buttons, cracked and scratched reservoir tube window, cracked reservoir tube lip, cracked display window, cracked case at display window corner and cracked battery tube threads.(b)(4).
|