• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Perforation (2001)
Event Date 07/09/2019
Event Type  Injury  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was unable to be reviewed as we were not able to obtain the serial number for the iabp associated with this complaint.The customer did not provide the model name/number or serial number of the iabp involved.The unit was put back into clinical use following the event, as there was no problem with the iabp.It was reported to be working well.The full name of the event site is (b)(6) national hospital.
 
Event Description
It was reported that during cardiosave intra-aortic balloon pump (iabp) therapy using a sensation plus 50cc intra-aortic balloon (iab) catheter, an aortic arch perforation occurred, presumed secondary to balloon-tip-induced trauma.There was no reported malfunction of the iabp.The facility is not attributing the injury to the iab catheter and believes the event is related to a medical training issue.The iab used in this event will be submitted under a separate mfg report#.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key8847079
MDR Text Key152854032
Report Number2249723-2019-01187
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 07/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0998-00-0800-XX
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/12/2019
Initial Date FDA Received07/31/2019
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
SENSATION PLUS 50CC BALLOON CATHETER
Patient Outcome(s) Life Threatening;
-
-