Model Number D002
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/07/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the device battery had allegedly prematurely depleted.The device was explanted and replaced to resolve the event.The patient was stable with no adverse consequences.
 
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, an engineering-level longevity prediction calculation was completed to assess the rate of battery depletion.Given the programmed parameters and other data stored within the memory of the device, the results of this calculation indicated that the actual rate of battery depletion fell within an acceptable range.Next, a series of diagnostic tests were conducted that verified the performance of pacing, sensing, (defibrillation), and recording functions.Having met the engineering longevity prediction, functionally passed all returned product testing, and with no further information to indicate a product performance issue, we have concluded that this device experienced normal battery depletion.
 
Event or Problem Description
It was reported that the device battery had allegedly prematurely depleted.The device was explanted and replaced to resolve the event.The patient was stable with no adverse consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORIGEN MINI ICD DR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key8847173
Report Number2124215-2019-14716
Device Sequence Number2730645
Product Code LWS
UDI-Device Identifier00802526533747
UDI-Public00802526533747
Combination Product (Y/N)N
Initial Reporter CountryGR
PMA/510(K) Number
P960040/S306
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/02/2020
Device Model NumberD002
Device Catalogue NumberD002
Device Lot Number115785
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/03/2019
Initial Date Received by Manufacturer 05/07/2019
Supplement Date Received by Manufacturer08/22/2019
Initial Report FDA Received Date07/31/2019
Supplement Report FDA Received Date11/03/2019
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age66 YR
-
-