| Catalog Number |
71675384 |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during surgery, the nail used was labeled as a 130 degree nail but it was actually a 125 degree nail.The nail was inserted into patient and drilled when doctor realized something didn't seem right.Doctor attempted to insert the lag screw with no success.The nail was removed.Correct nail was then inserted into patient without any issues or concerns.No injury or harm to patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated device was returned and evaluated.A visual inspection of the returned device could not confirm the stated failure mode.The device is damaged from the attempted implantation.The device was manufactured in 2019.A dimensional inspection confirmed the stated failure mode.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
|
| |
|