Catalog Number SEP8
Medical Device Problem Codes Fracture (1260); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a venacaval thrombectomy procedure using an indigo system separator 8 (sep8).During the procedure, the tip of the sep8 distal to the separator "bead" fractured and lodged into the inferior vena cava (ivc) wall of fibrotic clot along the wall.The "bead" was still attached to the sep8 and it was embedded into what appeared to be old fibrous clot/tissue.The cause of this could possibly be that the cat8 was being maneuvered forward into the target location and it pinched the tip of the wire against the vessel wall and the fibrous clot as the physician was pulling back on the sep8.The procedure was successfully completed by using a new sep8.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Please note that the following section was inadvertently missed on the initial mfr report and is being updated on this follow-up # 01 mfr report: 3005168196-2019-01478.Please note that the following section was incorrectly reported on the initial mfr report and is being corrected on this follow-up #01 mfr report: 3005168196-2019-01478.Please note that this complaint was submitted to the fda by the user facility with the following reference number: (b)(4).
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the inferior vena cava (ivc) using an indigo system separator 8 (sep8).During the procedure, the wire tip distal to the sep8 bulb broke off and lodged into the wall of fibrotic clot in the target vessel.The physician made a decision to leave the wire tip in the clot and remove the sep8 from the procedure.The procedure was then completed using a new sep8.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM SEPARATOR 8
Common Device Name
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8847324
Report Number3005168196-2019-01478
Device Sequence Number2499151
Product Code DXE
UDI-Device Identifier00814548017440
UDI-Public00814548017440
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K161523
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/02/2019
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/18/2021
Device Catalogue NumberSEP8
Device Lot NumberF84730
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 07/02/2019
Supplement Date Received by Manufacturer08/16/2019
Initial Report FDA Received Date07/31/2019
Supplement Report FDA Received Date10/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age54 YR
Date Report Sent to FDA01/01/2005
-
-