| Model Number |
V-175420 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/01/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A vyaire support representative went on-site to evaluate the suspect device.The support representative performed the work order and was able installed the supplied demand valve.After testing the new demand valve with different increase volumes, all maneuvers were carried out properly by the new demand valve and the issue was resolved.The representative checked and tested the device through calibrations and test measurements and returned the device to service specifications.
|
| |
|
Event or Problem Description
|
|
The customer reported during the masterscreen body/diffusion testing.The demand valve leaks.The customer reported after inspiration the patient experienced (nausea) sickness, gag reflex, and "overblown" lung.The customer reported the shutter does not open adequately after exposure time and the patients exhales next to the mouthpiece.The customer reported being unable to carry out measurements and reported the most likely cause of the issue is the demand valve assembly.The customer reported the patient left the hospital without any further action (no surgery, no treatment).The customer reported there is no patient injury associated with the event.
|
| |
|