| Model Number |
ZXT300 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2018.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zxt300 intraocular lens (iol) was implanted into the patients right eye.The iol was explanted as the patient needed a different toric power, and replaced with a non-johnson and johnson iol.The patient went from a multifocal, toric iol to a non-multifocal iol.There was no additional intervention performed and the patient is doing fine post-operatively.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation: yes, returned to manufacturer on: 7/30/2019.Device returned to manufacturer: yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturing site for evaluation.Visual inspection shows the product is damaged, most probably to make explant possible.It can be seen that the lens is contaminated, dust particles are present.This is expected as the lens was transported from controlled environment during manufacturing to a non-sterile, non-environmental controlled area.No further anomalies were found.Investigation of the return sample and the condition of the return sample do not suggest the damage was introduced during manufacturing.The complaint cannot be confirmed.Manufacturing records review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints have been received for this production order number.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|