| Model Number |
MMT-523RNAS |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/17/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A test reservoir was installed and did not lock in place due to completely broken reservoir tube lip.Unable to perform the displacement test due to completely broken reservoir tube lip.Unit had missing reservoir tube o-ring.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the reservoir compartment was cracked and chipped but reservoir was able to lock in place when inserted into insulin pump.The customer¿s blood glucose was unknown at the time of incident.The customer was advice the insulin pump will need to be replaced.Advice to discontinue use of the pump and revert to backup plan per health care professional instructions.The insulin pump will be returned for analysis.
|
| |
|