MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.7MM VA LCKNG SCREW SLF-TPNG WITH T8 STARDRIVE RECESS 12MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.7MM VA LCKNG SCREW SLF-TPNG WITH T8 STARDRIVE RECESS 12MM; SCREW, FIXATION, BONE Back to Search Results
Model Number 02.211.012
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, the customer received the package for variable angle locking screw self-tapping with t8 stardrive recess with no product.There was no patient involvement.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: a product investigation was completed: the implant was not returned, and the investigation was completed based on the supplied image.The image showed an empty package and is confirmed.However, we can¿t confirm that a manufacturing issue occurred without the device in hand for the following reasons: the image only shows three of the perforated seals and cannot be confirmed that the fourth side is still sealed as the important information document is covering the seal.A manufacturing record evaluation could not be performed as the lot number could not be determined from the image.No definitive root cause was able to be determined as the circumstances surrounding the event are unknown.During the investigation no unidentified product design/manufacturing issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7MM VA LCKNG SCREW SLF-TPNG WITH T8 STARDRIVE RECESS 12MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8850308
Report Number2939274-2019-59479
Device Sequence Number1400835
Product Code HWC
UDI-Device Identifier10886982053565
UDI-Public(01)10886982053565
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K100776
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number02.211.012
Device Catalogue Number02.211.012
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/05/2019
Supplement Date Received by Manufacturer08/29/2019
Initial Report FDA Received Date08/01/2019
Supplement Report FDA Received Date08/29/2019
Patient Sequence Number1
-
-