| Model Number |
DL-344-10-K |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/08/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The explanted lead was not returned to neuropace for analysis.Review of the rns neurostimulator ecogs and lead impedances confirmed that the ecog data were suggestive of a potential lead break.
|
| |
|
Event or Problem Description
|
|
During a routine clinic appointment, the rns neurostimulator was interrogated, however impedance measurements would not read for the left depth lead, suggestive of a lead break.Further review of recorded ecogs noted that a change in the signals from the lead was noted around (b)(6) 2018.The patient did not recall a specific trauma or fall on or around that date but noted that he has a tendency to fall out of bed during generalized tonic-clonic seizures and hit his head on a nearby nightstand.No programming changes were made on the affected lead during the office appointment.During a routine rns battery replacement procedure performed on (b)(6) 2019, the affected lead was explanted and a new depth lead was implanted and connected.No complications were reported during the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) the explanted lead was returned to neuropace for analysis.The investigation included visual inspection and electrical testing.Confirmed lead damage.We were unable to determine the cause of the damage to the lead coil.
|
| |
|
Event or Problem Description
|
|
The product was received and investigated by neuropace.
|
| |
|