• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DORNOCH ULTRA DUO HIGH FLUID CART NON-STERILE INSTRUCTIONS FOR USE; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DORNOCH ULTRA DUO HIGH FLUID CART NON-STERILE INSTRUCTIONS FOR USE; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Catalog Number ULDU500
Device Problem Inaccurate Information (4051)
Patient Problem No Patient Involvement (2645)
Event Date 07/19/2019
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Malfunctioning level sensor.The previous repair report for ultra duo flex fluid cart serial number (b)(4) was reviewed and noted no related non-conformances, requests for deviation or any other issues with the repair.The previous repair report review found no issues with the device after repair and all verifications, inspections and tests were successfully completed.The dhr review found no issues with the device and all verifications, inspections, and tests were successfully completed.On (b)(6) 2019, it was reported from (b)(6) that the cylinders are reading full, but are in fact empty.On (b)(6) 2019, a zimmer biomet certified service repair technician was contacted about the cart and dispatched to be at the site.On (b)(6) 2019, the technician arrived at the site and confirmed that the cart was giving incorrect fluid level readings and alarming.The technician replaced the level sensor (part #91584 and lot code #0029256) and then verified that the unit was functioning as intended.The technician then returned the unit to service without further incident.The device was tested, inspected, and repaired.The root cause for the cylinder reading incorrectly was due to a malfunctioning level sensor.The level sensor reads fluid levels in the cylinder, but can give erroneous readings when malfunctioning.The reported event was confirmed during inspection of the device and the device was noted to be functioning as intended after the level sensor was replaced.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.Product evaluated by external contractor.
 
Event Description
It was reported that the cylinders are reading full, but are in fact empty.No adverse events were reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRA DUO HIGH FLUID CART NON-STERILE INSTRUCTIONS FOR USE
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer (Section G)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key8851313
MDR Text Key207032177
Report Number0001954182-2019-00047
Device Sequence Number1
Product Code JCX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133786
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/01/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberULDU500
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/19/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-