| Model Number |
4248-9 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/03/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.Device not returned.
|
| |
|
Event or Problem Description
|
|
The customer reported inaccurate psi values.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: other, other text: the returned sensor was evaluated.External visual inspection showed no damage.The sponge pads on the electrodes were removed.Fibers from the pads and dried gel deposits were observed on the electrodes.Continuity testing was performed and an open circuit was observed across all the electrodes.The gel and fiber residue was removed and continuity testing was performed.Good continuity was observed when performing continuity testing between the electrodes and the connector; however, open continuity was observed across the r1, r2, and l2 electrodes.The sensor was returned used.Functional testing could not be performed since the sensor is intended for single-patient use; the integrity of the sensor (gel; contacts; etc) is broken once it is removed from patient use., corrected data: additional information- d4 model # updated from 4248 to 4248-9.
|
| |
|
Event or Problem Description
|
|
The customer reported inaccurate psi values.No patient impact or consequences were reported.
|
| |
|