| Catalog Number |
B1060-100 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Material Rupture (1546); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/12/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
G9: exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.The device was received.Investigation is not yet complete.A follow report will be submitted with all additional relevant information.Na.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a superficial femoral artery.A 6.0x100mm armada 35 balloon dilatation catheter (bdc) was pressurized to an unspecified pressure and the balloon ruptured.During removal of the bdc, there was resistance noted with the anatomy and the balloon separated from the catheter.It was reported that the separated balloon was removed but the method is unknown.A new unspecified bdc was used to successfully complete the procedure.There was no adverse patient sequela reported and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
A visual inspection was performed on the returned device and the reported balloon rupture and balloon separation were confirmed.The reported difficulty to remove was unable to be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.There was no damage noted to the balloon catheter during the inspection prior to use which suggests a product quality issue with respect to manufacture, design or labeling did not contribute to the reported difficulties.The investigation determined the reported balloon rupture, difficulty to remove and balloon separation and noted damages appear to be related to the operational circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling; therefore, no corrective action is required.Exemption number e2019001.
|
| |
|