• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number B1060-100
Medical Device Problem Codes Difficult to Remove (1528); Material Rupture (1546); Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/12/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
G9: exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.The device was received.Investigation is not yet complete.A follow report will be submitted with all additional relevant information.Na.
 
Event or Problem Description
It was reported that the procedure was performed to treat a superficial femoral artery.A 6.0x100mm armada 35 balloon dilatation catheter (bdc) was pressurized to an unspecified pressure and the balloon ruptured.During removal of the bdc, there was resistance noted with the anatomy and the balloon separated from the catheter.It was reported that the separated balloon was removed but the method is unknown.A new unspecified bdc was used to successfully complete the procedure.There was no adverse patient sequela reported and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
A visual inspection was performed on the returned device and the reported balloon rupture and balloon separation were confirmed.The reported difficulty to remove was unable to be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.There was no damage noted to the balloon catheter during the inspection prior to use which suggests a product quality issue with respect to manufacture, design or labeling did not contribute to the reported difficulties.The investigation determined the reported balloon rupture, difficulty to remove and balloon separation and noted damages appear to be related to the operational circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling; therefore, no corrective action is required.Exemption number e2019001.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key8852466
Report Number2024168-2019-10509
Device Sequence Number16520176
Product Code LIT
UDI-Device Identifier08717648154720
UDI-Public08717648154720
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2021
Device Catalogue NumberB1060-100
Device Lot Number80706G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/31/2019
Initial Date Received by Manufacturer 07/12/2019
Supplement Date Received by Manufacturer08/16/2019
Initial Report FDA Received Date08/01/2019
Supplement Report FDA Received Date09/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
-
-