Model Number 520-01-250
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 07/02/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported as an infection.The previous surgery and the revision detailed in this investigation occurred 6.8 yrs.Apart.The healthcare professional indicated there was a serious risk to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhrs) show that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the infection.The device and its applicable concomitant devices were within their respective expiration dates at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The device was verified to have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery.The root cause of this complaint was reported as an infection.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the patient's infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection or inhibited the patient's immune system.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - patient presented to surgeon's office with symptoms of infection.Surgeon opted to irrigation & drainage and remove all implants.Cemented spacer was implanted temporarily to combat infections.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Common Device Name
TURON PEGGED GLENOID IMPLANT, SIZE 50
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin,tx 78758-5445
MDR Report Key8852928
Report Number1644408-2019-00703
Device Sequence Number16826975
Product Code KWS
UDI-Device Identifier00888912024976
UDI-Public(01)00888912024976
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K123982
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2016
Device Model Number520-01-250
Device Catalogue Number520-01-250
Device Lot Number193G1007
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/02/2019
Initial Report FDA Received Date08/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age87 YR
Patient Weight95
-
-