| Model Number |
DL-344-10-K |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The explanted lead was not returned to neuropace for analysis.However review of the rns neurostimulator ecog and lead impedance data confirmed that the ecog data were suggestive of a potential lead break.
|
| |
|
Event or Problem Description
|
|
Review of the rns neurostimulator ecog data indicated a lead artifact on the right depth lead.Therapy and detection were disabled on the affected lead.No information regarding seizure-related trauma or other potential causal factors resulting in the lead break were provided by the treating center.On (b)(6) 2018 the affected lead was explanted and a new lead was implanted.No complications were reported during the procedure.
|
| |
|