Model Number VNMC3737C90TU
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Intimal Dissection (1333); Occlusion (1984)
Date of Event 07/16/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A valiant navion stent graft system was implanted in a patient for the endovascular treatment of a thoracic aortic dissection, from the lsa to the abdominal aorta.It was reported that during the implant procedure, a 37/90 valiant navion stent graft was implanted at the lcc and was then extended with a 43/37/200 valiant navion, with perfect overlap.It was reported that the physicians were then concerned that the 37/90 valiant navion was encroaching on the lcc, and they decided to place an 8/29 balloon-expandable bare stent for access.It was reported that on the day after the procedure, the patient presented emergently with an rtad.An open repair was performed as treatment.As per the physician, the cause of the event could not be determined.No additional clinical sequelae were reported and the patient is fine.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT NAVION
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key8854083
Report Number9612164-2019-03190
Device Sequence Number1904454
Product Code MIH
UDI-Device Identifier00763000101053
UDI-Public00763000101053
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P100040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/24/2021
Device Model NumberVNMC3737C90TU
Device Catalogue NumberVNMC3737C90TU
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/18/2019
Supplement Date Received by Manufacturer07/18/2019
Initial Report FDA Received Date08/02/2019
Supplement Report FDA Received Date08/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
-
-