| Model Number |
8881501160 |
| Medical Device Problem Code |
Nonstandard Device (1420)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/25/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The incident sample has been requested but to date has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
|
| |
|
Event or Problem Description
|
|
The customer reported that the plunger is sticky and hard to pull back.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.The manufacturing site received eleven sealed package samples for evaluation.A complete investigation was performed.Visual inspection to the quality inspection standard was conducted and no defects were identified.The lab performed plunger activation force testing on the eleven samples received and all samples were found to be within specification.The reported condition could not be confirmed.A root cause could not be determined as the reported issue was not confirmed.A corrective action is not applicable at this time.Functional testing and visual inspections are being performed according to our current quality standards and inspection procedures.This complaint will be used for qa tracking and trending purposes.
|
| |
|