• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ADVANCE 35 LP LOW PROFILE BALLOON CATHETER; LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ADVANCE 35 LP LOW PROFILE BALLOON CATHETER; LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number G35536
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/25/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported prior to a leg angiogram procedure involving a patient of unknown gender and age, a hole was observed upon opening the advance 35 lp low profile balloon catheter.A 6 fr.Sheath was going to be used; however, the device never made patient contact, nor was it inflated to nominal or burst pressure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Investigation/evaluation: reviews of the complaint history, device history record, instructions for use (ifu), and quality control procedures of the device were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Furthermore, a review of the quality control procedures was conducted, and no gaps were discovered.Moreover, an ifu is provided with this device, which states that the device is indicated for percutaneous transluminal angioplasty (pta) of lesions in peripheral arteries; including iliac, renal, popliteal, infrapopliteal, femoral and iliofemoral; and obstructive lesions of native or synthetic arteriovenous dialysis fistulae.The document also instructs that upon removal, the product is to be inspected to ensure no damage has occurred.Based on the information provided and no product returned, investigation has concluded that a root cause for this event could not be established.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 35 LP LOW PROFILE BALLOON CATHETER
Common Device Name
LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key8854868
Report Number1820334-2019-01912
Device Sequence Number11336583
Product Code LIT
UDI-Device Identifier10827002355368
UDI-Public(01)10827002355368(17)220503(10)9712899
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K132020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/03/2022
Device Model NumberG35536
Device Catalogue NumberPTA5-35-135-6-20.0
Device Lot Number9712899
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/25/2019
Supplement Date Received by Manufacturer10/09/2019
Initial Report FDA Received Date08/02/2019
Supplement Report FDA Received Date10/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-