• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEVICOR MEDICAL PRODUCTS, INC. MAMMOMARK; BIOPSY SITE IDENTIFIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEVICOR MEDICAL PRODUCTS, INC. MAMMOMARK; BIOPSY SITE IDENTIFIER Back to Search Results
Model Number MAM3001
Device Problem Device Damaged by Another Device (2915)
Patient Problem Foreign Body In Patient (2687)
Event Date 07/23/2019
Event Type  Injury  
Manufacturer Narrative
The mammomark biopsy site identifier is intended for use after an open surgical or percutaneous breast biopsy procedure to the biopsy site.The device is not available for analysis, which precludes a full investigation and analysis of the root cause.Our mammotome vacuum assisted biopsy probes contain extremely sharp edges along the aperture opening to effectively excise tissue.Removing the marker delivery system separately from the probe aperture once the spring is exposed creates the possibility of it catching on one of these edges.As a mitigation step to address this risk, we provide instructions within the instructions for use: directions: remove the delivery system and mammotome probe together as a single unit from the site, properly dispose and obtain images to confirm marker placement.
 
Event Description
Devicor medical products, inc.Received a report from user facility, stating, tip shear occurred during stereo biopsy.Doctor left in the breast because they are waiting to see if patient will go to surgery once pathology report comes back.Doctor pulled marker out of the probe instead of removing the probe/marker as one unit.This has been documented in our system as record #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAMMOMARK
Type of Device
BIOPSY SITE IDENTIFIER
Manufacturer (Section D)
DEVICOR MEDICAL PRODUCTS, INC.
300 e. business way
fifth floor
cincinnati OH 45241
Manufacturer (Section G)
DEVICOR MEDICAL PRODUCTS DE MEXICO
sor juana ines de la cruz
#20152 4-b, parque industrial
tijuana,
Manufacturer Contact
jill burnett
300 e. business way
fifth floor
cincinnati, OH 45241
MDR Report Key8854930
MDR Text Key153088843
Report Number3008492462-2019-00039
Device Sequence Number1
Product Code NEU
UDI-Device Identifier10841911101595
UDI-Public(01)10841911101595(17)200221(10)F11833475D
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082278
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/02/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/21/2020
Device Model NumberMAM3001
Device Catalogue NumberMAM3001
Device Lot NumberF11833475D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/23/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/20/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-