The physician reported that a patient was having therapy related pain.There was no high lead impedance, and per the physician the patient's x-ray looked normal; however, the physician still believed that there may be something wrong with the generator.The sales representative visited the patient in the hospital to check the device.The sales representative reported that the system diagnostics were within normal limits.Clinic notes were later received confirming that the patient had been referred for generator/lead revision to upgrade to a m1000 sentiva generator.The clinic notes were reviewed and mentioned that the patient continues to have intermittent chest pains.The physician noted that the plan for the reported pain was to refer the patient to the surgeon for replacement.No other relevant information has been received to date.
|
The explanted lead and generator were received by the manufacturer.The generator and lead were explanted and returned due to reported pain and painful stimulation.A comprehensive automated electrical evaluation of the generator showed that the pulse generator performed according to functional specifications.The device output signal was monitored for more than 24-hrs, while the pulse generator was placed in a simulated body temperature environment.Results showed no signs of variation in the pulse generator¿s output signal and demonstrated that the device provided the expected level of output current for the entire monitoring period.Additionally, it was confirmed that the septum was not cored, thus eliminating the possibility of a potential unintended electrical current path through body fluids.There was no performance, or any other type of adverse conditions found with the pulse generator.The condition of the returned lead portion is consistent with conditions that typically exist following an explant procedure.Other than a slice mark and cut out hole found on the outer silicone tubing, most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing, there were no obvious anomalies found.What appeared to be white deposits were observed in various locations.Energy dispersion spectroscopy confirmed that these deposits contained silicon, phosphorus, sodium, magnesium and calcium.The setscrew marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portion were performed, during the visual analysis, and no discontinuities were identified.Based on the findings in the product analysis lab, there is no evidence to suggest an anomaly with the returned portion of the device which may have contributed to the stated complaints.Note that since the electrode array section was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.No other relevant information has been received to date.
|