• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number X SERIES
Medical Device Problem Code Unable to Obtain Readings (1516)
Health Effect - Clinical Code Death (1802)
Date of Event 07/20/2019
Type of Reportable Event Death
Additional Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event or Problem Description
Complainant alleged that while attempting to treat a patient in cardiac arrest (age & gender unknown),the device was unable to obtain an ecg signal via electrode pads.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired.
 
Additional Manufacturer Narrative
The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a faulty mfc-to-cprd connector.The device was recertified and returned to the customer.The customer received a replacement mfc-to-cprd connector.Analysis for reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key8855818
Report Number1220908-2019-02254
Device Sequence Number2499072
Product Code DRT
UDI-Device Identifier00847946006372
UDI-Public00847946006372
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K112432/P160
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberX SERIES
Device Catalogue NumberX SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/31/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/24/2019
Supplement Date Received by Manufacturer07/24/2019
Initial Report FDA Received Date08/02/2019
Supplement Report FDA Received Date09/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
-
-