This report is submitted beyond 30-calendar days from allergan¿s receipt due to the exemption transition period.A review of the device history record has been completed.No deviations or non-conformances noted.The reason for reoperation is rupture.Device evaluation: visual analysis of the returned device identified; wear abrasion, the weight the device within specification, crease fold, deformation and cloudy color in the gel observed after autoclave cycle.Based on the device analysis the final assessment is: no were observed openings on the device or issues related with the complaint (rupture).Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
|