• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY ACROBAT® 2 CALIBRATED TIP WIRE GUIDE; OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY ACROBAT® 2 CALIBRATED TIP WIRE GUIDE; OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Model Number G47617
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/09/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products : mtw ercp catheter, 215 cm, 0.025", unknown reference part number.Investigation evaluation: our laboratory evaluation of the product said to be involved confirmed the report.The tip of the device was bent at approximately 3 cm from the distal end.Approximately 11.8 cm to 13.8 cm from the distal end is a section of bare core wire with intact coating attached at either end.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.The device history records for the wire guide lot was reviewed.The wire guide device history record contains a nonconformance that could potentially be related to wire guide damage.The device goes through different inspections prior to leaving the facility.These inspections would have removed any nonconforming devices prior to distribution.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.If additional pressure is applied to the wire guide and/or accessory device(s) while moving the wire guide inside the accessory device(s), this could contribute to wire guide coating damage.Prior to distribution, all acrobat® 2 calibrated tip wire guides are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook acrobat® 2 calibrated tip wire guide.Procedure for replacing a bile duct stent was performed.Wire-guided-cannulation was performed from the side of the stent with the wire guide and another manufacturer's ercp catheter.The ercp catheter did not have a metal tip.After that, the ercp stent [catheter] was removed and the wire guide remained at the side of the stent.Then, the user attempted to remove the stent but coating material of the wire guide peeled off during removal of the stent.Another wire guide was used instead [lost wire guide access] [subject of this report].There have been no adverse effects to the patient reported.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACROBAT® 2 CALIBRATED TIP WIRE GUIDE
Common Device Name
OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
MDR Report Key8857486
Report Number1037905-2019-00438
Device Sequence Number2397244
Product Code OCY
UDI-Device Identifier00827002476172
UDI-Public(01)00827002476172(17)220411(10)W4202512
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/11/2022
Device Model NumberG47617
Device Catalogue NumberAWG2-35-450-A
Device Lot NumberW4202512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2019
Initial Report FDA Received Date08/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OLYMPUS JF-260V ENDOSCOPE
-
-