| Catalog Number |
112280 |
| Medical Device Problem Codes |
Output Problem (3005); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As part of the normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
During a mako tka the surgeon had difficult maintaining a registration on the tibia and femur.On inspection the array itself was wobbly at the centre attachment.Surgical delay of approximately 1 hour.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow-up #1 and final report submitted to update sections d.3, g.1, g.4, g.7, h.2, h.3, h.6, h.10 and h.11 based on the results of investigation.Reported event: it was reported that "during a mako tka the surgeon had difficult maintaining a registration on the tibia and femur.On inspection the array itself was wobbly at the centre attachment." the event was not confirmed.Method & results: -device evaluation and results: not performed as no items were returned.-product history review: review of the device history records indicates 100 device(s) were manufactured and accepted into final stock on 14may2018 with no reported discrepancies.-complaint history review: a review of complaints in catsweb and trackwise related to p/n 112280, lot 19160218 shows 00 additional complaint(s) related to the failure in this investigation.Conclusion: the exact cause of the event could not be determined because the product was not received.No further investigation for this event is possible at this time as no devices and / or insufficient information was received by stryker orthopaedics.If devices and / or additional information become available, this investigation will be reopened.Product surveillance will continue to monitor for trends corrective action/preventive action: no action is required at this time as there is no confirmed failure mode.H3 other text : product was not available for evaluation.
|
| |
|
Event or Problem Description
|
|
During a mako tka the surgeon had difficult maintaining a registration on the tibia and femur.On inspection the array itself was wobbly at the centre attachment.Surgical delay of approximately 1 hour.
|
| |
|