This report is submitted beyond 30-calendar days from allergan¿s receipt due to the exemption transition period.A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified: opening, weight to the spec.A microscopic analysis was performed which identify a striated edge opening, consist with use of a surgical tool.Based on the device analysis the final assessment is: a striated edge opening on radius side due to surgical damage.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.This is a known potential adverse event addressed in the product labeling.
|