| Model Number |
CN-830-L |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Information (3190)
|
| Date of Event |
07/09/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, halfway through the surgery, the thread broke from the needle.
|
| |
|
Event or Problem Description
|
|
According to the reporter, halfway through a hernia repair, the thread broke from the needle.Another device was used to complete the case.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) received twenty devices.A tactile and microscopic inspection of the sutures was performed with no abnor malities.A needle attachment test was performed on the sealed samples, and the results met the specifications for this product.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|