Model Number 18780
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Information (3190)
Date of Event 07/05/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the pilot valve of the eleven tubes deflated in a short period.It was reported that the tube lost pressure after 2 to 3 days on average and was tested with cufometer with pressure not stable.After taking out of the patient, the tube was tested and there was no leakage in the cuff.Reintubation was required to prevent a permanent impairment of a function.
 
Additional Manufacturer Narrative
New information received.This report will now be captured under mfr report number: 2936999-2019-00536 new pe (b)(4).Any additional information received in the future will be captured and submitted under the report number 2936999-2019-00536 new pe (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MALLINCKRODT
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
MMJ SA DE CV(USD)
ave henequen no 1181 desarrol
ciudad juarez,mx 32590
MX  32590
MDR Report Key8858766
Report Number2936999-2019-00509
Device Sequence Number10891086
Product Code BTR
UDI-Device Identifier50884522007618
UDI-Public50884522007618
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K090352
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/22/2023
Device Model Number18780
Device Catalogue Number18780
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/17/2019
Supplement Date Received by Manufacturer07/17/2019
Initial Report FDA Received Date08/05/2019
Supplement Report FDA Received Date08/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1