• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPSMEDICAL SUPPLY CORP. STEAM TAPE; INDICATOR, PHYSICAL/CHEMICAL STERILIZATION PROCESS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPSMEDICAL SUPPLY CORP. STEAM TAPE; INDICATOR, PHYSICAL/CHEMICAL STERILIZATION PROCESS Back to Search Results
Model Number ST-036
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 05/21/2019
Type of Reportable Event Malfunction
Event or Problem Description
The tape for sterile processing did not turn colors after sterilizing the trays as intended, to confirm sterility.Trays were re-opened and reprocessed as a precaution.All machine parameters had been met.Integrators inside the trays had passed.Some loads were processed with biological indicators, which had also passed.We pulled the lot number of the tape that we believed to be affected and ran a test.Tape from 2 different rolls of the same lot number had different results.1 was fine, 1 was not.We thought that it was just 1 bad roll.This morning there were several more trays that the tape did not look as dark, so we pulled and reprocessed those as well.We ran another test and we were not able to recreate the issue.We have now pulled all rolls with that lot number.This error is a good catch.It did not reach the patient or cause any harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEAM TAPE
Common Device Name
INDICATOR, PHYSICAL/CHEMICAL STERILIZATION PROCESS
Manufacturer (Section D)
SPSMEDICAL SUPPLY CORP.
6789 west henrietta rd.
rush NY 14543
MDR Report Key8860229
Report Number8860229
Device Sequence Number11284495
Product Code JOJ
UDI-Device Identifier10732224730193
UDI-Public(01)10732224730193(10)9017(17)220131
Combination Product (Y/N)N
Initial Reporter StateNY
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/26/2019
Report Date (Section F) 07/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberST-036
Device Catalogue NumberST-036
Device Lot Number9017
Was the Report Sent to FDA? Yes
Event Location Hospital
Date Report to Manufacturer08/05/2019
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA07/15/2019
-
-