| Model Number |
CD3231-40Q |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/22/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Interrogation of the device revealed it was above eri when received.Based on this information, the device was found to communicate appropriately with a merlin programmer and has not reached the elective replacement indicator (eri).The device is included in the premature battery depletion with implantable cardioverter defibrillator advisory notice issued by st.Jude medical on 11 october 2016.
|
| |
|
Event or Problem Description
|
|
Following the advisory for premature battery depletion with implantable cardioverter defibrillator, although there was no eri alert or allegation of premature battery depletion, the device was explanted prophylactically. there were no patient consequences.
|
| |
|