| Catalog Number |
0064599 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2019 immucor technical support used a remote electronic connection method to assess files on galileo neo instrument serial number (b)(4); roi's were acceptable and no errors were seen during times of testing.On (b)(6) 2019 an immucor field service technician inspected galileo neo instrument serial number (b)(4) at the customer facility.The technician found the instrument operational and performed a successful unexpected reaction checklist.Found instrument to be fully operational and to be operating in specification.Customer was informed that unable to rule out the possibility that the original vials of anti-d (monoclonal blend) series 4 and anti-d (monoclonal blend) series were compromised.The immucor internal report record number for this report is (b)(4).
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019 a customer reported rh discrepancies when testing donor samples on the galileo neo instrument.
|
| |
|
Search Alerts/Recalls
|