| Catalog Number |
08K28-03 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/05/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Correction/removal number: 3002809144-06/4/19-006-r.A product recall letter was issued to all architect bnp customers who have received calibrator or control lots still within dating.The letter informs the customer of the issue regarding a time dependent stability issue of the architect bnp calibrators and controls that may lead to controls out of range and shift in control and patient results.The letter informs the customer all architect bnp calibrators and controls will have shortened expiration dates of 165 days from the date of manufacture.The letter instructs the customer to discontinue use of the lots which are beyond the 165 day expiration and destroy any remaining inventory.The cause of the shift is instability of the architect bnp calibrators and controls.
|
| |
|
Event or Problem Description
|
|
The customer obtained falsely elevated and failing (b)(6) survey results while using the architect bnp calibrator lot 44k82018.Sample (b)(6) generated a result of 1319.6 with an expected range of 789 to 1239.6 and sample (b)(6) generated a result of >5000 with an expected range of 3579.0 to 4928.7.There was no report of incorrect patient results or impact to patient management.
|
| |
|