| Catalog Number |
120-600 |
| Medical Device Problem Codes |
Material Rupture (1546); Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
04/08/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This report is submitted beyond 30-calendar days from allergan¿s receipt due to the exemption transition period.A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Reason for reoperation was due to capsular contracture, baker grade iv and a rupture.Device evaluation: visual analysis of the returned device identified: creases, bubbles and deformation.Cloudy after autoclave disinfection process in the gel.A weight test of the device was verified and the device was within specification.Based on the device analysis the final assessment is: no were observed openings on the device or issues related with the complaint (rupture).This is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side capsular contracture baker grade iv and a rupture.Device has been explanted.
|
| |
|