| Model Number |
M00558430 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Investigation result: a visual examination of the returned complaint device revealed that the balloon was burst near the proximal bond of the balloon.Functional evaluation could not be performed due to the condition of the returned device.This failure is likely due to factors or conditions related to procedure during the use of the device that could have affected its performance and its intended purpose.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre wire guided dilatation balloon was used during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed in the major duodenal papilla on (b)(6) 2018.According to the complainant, during the procedure, it was noted that the balloon was leaking upon inflation.The procedure was completed with another cre wire guided dilatation balloon.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.Investigation results revealed that the balloon was burst; therefore, this is now an mdr reportable event.
|
| |
|