| Catalog Number |
0279351250 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the probe was overheating and produced smoke.A replacement was available.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: heated up and smoked.Probable root cause: heat can result from insufficient suction which can result from: inadequate hospital suction, leak, bad connection to hospital suction line, clogging caused by suction path blockage, lower electrode mass due to variation in electrode thickness or opening cut.Use error: the reported failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the probe was overheating and produced smoke.A replacement was available.
|
| |
|