| Model Number |
PCB00 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
06/14/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age/date of birth: unknown, information not provided.Gender/sex: unknown, information not provided.If implanted, if explanted, give date: not applicable, as lens was not implanted.(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that preloaded pcb00 device did not have an intraocular lens in the cartridge when surgeon went to implant.There was patient contact with the cartridge.Reportedly sutures were required during this procedure.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation; returned to manufacturer on: 8/1/2019.Device evaluation: the product was returned to the manufacturing site for evaluation.The returned pcb00 device was received in a plastic bag.The plunger was observed in advanced position.The plunger was gently pulled back and found locked and functional.The pcb00 device was observed under microscope and traces of viscoelastic and/or balanced salt solution were observed at the tip of the cartridge.Dfu z311134p dfu, tecnis itec preload pcb00, 1st revision a states to completely fill the viewing window of the pcb00 with ovd.The lens was observed stuck and overridden at the cartridge.There were no damages observed on the assembly.The reported complaint issue was not verified.No product quality deficiency was identified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints related to this production order was conducted in the complaint system.The search revealed no additional investigation requests (ir) received for this production order.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|