|
Suspected device evaluated by ok biotech and calculated that the meter operated within specifications.We tested the standby current of returned meter, the result was 0.9¿a.The criteria is <55¿a.Pass.Meter setting, audio and all buttons function are ok.We tested the suspected meter with in house control solution and returned strips (strip lot number: d171117-2, only (b)(4) pcs).The control solution tests for level low were 84/87 mg/dl, for level high were 289/279 mg/dl.The request control solution ranges are: level low 40~90 mg/dl; level high 220~330 mg/dl.All results were within the acceptance range.Pass.However we found the desiccant inside the strip bottle had color changed.We tested the retain strips from our warehouse (strip lot number: d171117-2).The control solution tests for level low were 65/70 mg/dl; for level high were 261/252 mg/dl.The request control solution ranges are: level low 40~90 mg/dl; level high 220~330 mg/dl.All results were within the acceptance range.Pass.Because the desiccant color changed from patient's strips, indicate the strips might get moisture.Patient used those strips to test blood might caused or contributed to higher readings.User storage or operation issue.
|