SMITH & NEPHEW, INC. CHONDRAL PICK HANDLE ASSY; ARTHROSCOPE Back to Search Results
Catalog Number 014800
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/09/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during surgery, the chondral pick broke when hit with a hammer, as indicated by technique.The tip was removed by the physician.The procedure was finished with a plaster instead to successfully complete the surgery.Neither surgical delay nor patient injuries have been indicated.
 
Additional Manufacturer Narrative
One 014800 chondral pick handle assembly was used for treatment but was not returned for evaluation.This is a nine year old reusable instrument component.The symptom is aligned with having blunt force applied but due to product unavailability, the complaint could not be ultimately confirmed.Definitive conclusions, accurate investigation and evaluation were limited without evaluation of physical product.If objective evidence, relevant information, packaging or product becomes available to assist with evaluation, the complaint will certainly be revisited.Factors that may affect device performance include: device ability, surgical ability, implant location and tissue condition.Review of instruction for use (ifu) documentation confirms instructions, precautionary statements and recommendations for proper use of product.Instrument tips may inadvertently be twisted or caught on other instruments or the basket during cleaning.This can lead to compromised strength.Per ifu: ¿these instruments are designed for repeated use with proper care and handling.As with any surgical instrument, careful attention should be made to assure that excessive force is not placed on the instrument.Excessive forces applied to the instrument can result in the instrument¿s failure.Do not use these instruments as levers for manipulating hard tissue or bone.Excessive force should not be applied to the instrument when manipulating soft tissue, bone, or hard objects.Misuse of these instruments may result in bent distal tips or jaws; and dull or uneven cutting edges.¿ product met specifications upon release to distribution.
 
Additional Manufacturer Narrative
One 014800 chrondral pick handle assembly used for treatment, was returned for evaluation.The complaint stated: ¿the chondral pick broke when hit with a hammer, as indicated by technique.¿ this was a nearly nine year old reusable pick.The incident is aligned with repeat instances of hammering.Tapping is recommended.The product has received high impact.Material has been crushed, dented and rolled from force.Inspection prior to each use is recommended.Per instructions for use ¿these instruments are designed for repeated use with proper care and handling.As with any surgical instrument, careful attention should be made to assure that excessive force is not placed on the instrument.Excessive forces applied to the instrument can result in the instrument¿s failure.Do not use these instruments as levers for manipulating hard tissue or bone.Excessive force should not be applied to the instrument when manipulating soft tissue, bone, or hard objects.Misuse of these instruments may result in bent distal tips or jaws; and dull or uneven cutting edges.¿ product met specifications upon release to distribution.No root cause related to the manufacturing of this product was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHONDRAL PICK HANDLE ASSY
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key8863976
Report Number1219602-2019-00917
Device Sequence Number2742579
Product Code HRX
UDI-Device Identifier03596010026408
UDI-Public03596010026408
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K940985
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/04/2020
Device Catalogue Number014800
Device Lot Number50367646
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2019
Initial Date Received by Manufacturer 07/10/2019
Supplement Date Received by Manufacturer08/27/2019
10/30/2019
Initial Report FDA Received Date08/06/2019
Supplement Report FDA Received Date08/28/2019
10/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age32 YR
-
-