• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR SILTEX ROUND MODERATE PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR SILTEX ROUND MODERATE PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Catalog Number 3542615
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 07/16/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Concomitant products: left-mentor siltex round moderate profile 150cc saline breast implant, catalog number: 3542615, lot number: 195595.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) year-old caucasian female patient underwent a primary breast augmentation with mentor siltex round moderate profile 150cc saline breast implants which the right side deflated after implantation.The issue was confirmed by the physician.As a result patient underwent bilateral removal and replacements with mentor gel device on (b)(6) 2019.
 
Additional Manufacturer Narrative
Device received on (b)(6) 2019.Device evaluation completed on (b)(6) 2019 during evaluation of the sample it was noticed a tear in the posterior view, measuring approximately 0.5 cm.Microscopic examination was performed and no evidence of instrument damage was observed.No other anomalies were observed.Deflation is a known complication associated with mentor mammary prostheses.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet.It should be noted that as part of our quality process, each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Based on the information reported and/or the product investigation, there is no evidence that the issue is related with the manufacturing process.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SILTEX ROUND MODERATE PROFILE
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key8864375
Report Number1645337-2019-16489
Device Sequence Number19592825
Product Code FWM
UDI-Device Identifier00081317001720
UDI-Public00081317001720
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2003
Device Catalogue Number3542615
Device Lot Number195595
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/16/2019
Initial Date Received by Manufacturer 07/18/2019
Supplement Date Received by Manufacturer08/16/2019
Initial Report FDA Received Date08/06/2019
Supplement Report FDA Received Date09/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
-
-