It was reported that halfway through a knee arthroscopy, the pump started running at a much higher rate than indicated and never achieved the desired pressure on the display.The pump was set on 35 and there was no alarm when this was occurring.The case was stopped due to the excessive fluid and the patient developed compartment syndrome.An additional incision was made as a fasciotomy was performed to reduce the pressure.The case was aborted and the procedure will be rescheduled.The rep noted the reservoir was halfway full of fluid before the case started.He warned the tubing needed to be set up again but the nurse stated it was set up correctly as it always has in the past.The doctor went on to start the case despite the warning.Pump setting was 35.The planned procedure was a knee arthroscopy chondroplasty and possible oats.
|