| Model Number |
MT20649 |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that error icon displayed occurred.No product or data was provided for evaluation.Confirmation of the allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that error icon displayed occurred.The product was evaluated.An external visual inspection was performed and passed.A receiver charge test was performed and passed.A receiver boot test was performed and failed due to it would not boot.Functional tests were performed and passed.The log was downloaded and reviewed finding error related to the complaint.The allegation was confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Product was received but is pending evaluation.A follow up report will be submitted upon completion.No injury or medical intervention was reported.
|
| |
|
Search Alerts/Recalls
|