| Model Number |
9-ITV10F45/80 |
| Medical Device Problem Code |
Material Perforation (2205)
|
| Health Effect - Clinical Codes |
Perforation (2001); Pericardial Effusion (3271)
|
| Date of Event |
07/17/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information regarding this event has been requested.The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, a 16mm amplatzer muscular vsd occluder was selected for implant.The procedure was complex as the patient's defect was very apical.There was also the presence of a pseudo-aneurysm located on the lateral wall of the vd (ventricle defect) and a litany of trabeculations.As the physician was getting ready to deploy the device and a pericardial effusion on the lateral side of the vd was noted.The physician quickly deployed the occluder and punctured the pericardium.220cc of blood was extracted and after 125 minutes, a pericardial drain was set in place and the bleeding stopped.The procedure took longer than anticipated.The patient is reported to be in stable condition.Manufacturer report number: 2135147-2019-00223.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: upon review, the delivery system should not have been submitted as a medical device report (mdr) as the event did not indicate a malfunction caused a serious event.
|
| |
|