| Model Number |
179732645 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Expedium 5.5 titanium screws.1797-32-645 - 4 pieces.Reason for contact: intraoperatively, while tightening the reduction screw with a nut-screw, the tuning forks of the two screws prematurely broke off.The incident was reported to the operating surgeon.Date / time of the incident: (b)(6) 2019 13:00.Date / time when j & j employee became aware of the incident: july 15, 2019 13:00.Photos of products attached.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Udi: (b)(4).Report type: corrected data.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was not returned for evaluation.A review of the device history record could not be performed as the lot number is unknown.Without the return of the device, we are unable to confirm the reported issue or identify the root cause.This complaint file will be closed with no further action required.Should more information and/or the sample be provided at a later time, this complaint will be reopened and device evaluated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Expedium 5.5 titanium screws: 1797-32-645 - 4 pieces.Reason for contact: intraoperatively, while tightening the reduction screw with a nut-screw, the tuning forks of the two screws prematurely broke off.The incident was reported to the operating surgeon.Date / time of the incident: (b)(6) 2019 13:00.Date / time when j & j employee became aware of the incident: july 15, 2019 13:00.Concomitant device reported: si polyaxl ext tab 6 x 45mm (part# 179732645, lot# unknown, quantity 2).This complaint involves two (2) devices.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).
|
| |
|