• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION, INC BD TRAY SPN WHIT27G3.5 L/B-D/E; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION, INC BD TRAY SPN WHIT27G3.5 L/B-D/E; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number 405699
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 07/22/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Initial reporter phone #: (b)(6).Investigation summary: three photos were provided for evaluation.Vial appears as intended with no obvious defects.Fluid appears to be clear.No avenue to evaluate drug potency without the physical sample.Failure mode could not be confirmed.Over a period of more than 10 years, no test result from samples returned due to ineffective anesthesia has ever failed to meet specifications.The causes of ineffective anesthesia have been well documented in clinical literature for many years.Whereas a strong history of testing has supported that drug potency is not the cause, the actual cause may not always be ascertained.Bd has a long history of test results to support that drug potency issues have not been the cause of ineffective anesthesia events reported to bd through the complaint system.As a preventive action, the vendor of the applicable drug component (b)(4) will be notified of the reported failure mode from the customer.Likewise, the complaint will be entered into the complaint management system and will be tracked & trended for future occurrences and reviewed/investigated through the quality data analysis (qda) process if an adverse trend is identified.A device history record review of all applicable manufacturing records for lot 0001270921 did not identify any issues that may have contributed to the reported failure mode.Investigation conclusion: a complaint history check was performed and this is the 4th related complaint reported with the defect/condition of anesthesia ineffective with lot #1270921 regarding item #405699 related complaints: (b)(4).A device history record review of all applicable manufacturing records for lot 0001270921 did not identify any issues that may have contributed to the reported failure mode.Root cause description: based on the complaint investigation, a probable root cause could not be identified for the reported failure mode.The investigation identified a previous capa (capa (b)(4)) performed by the bd anesthesia quality group that specifically investigated the ¿ineffective anesthesia¿ failure mode.Refer to capa (b)(4) for additional information regarding the outcome of the investigation.Likewise, the investigation identified a summary for previously investigated complaints (pic: (b)(4)) which provides rationale and information regarding the potential ineffective anesthesia outcome during the use of an anesthesia product code.Based on a review of all of these sources, the current complaint investigation could not identify a definitive root cause nor could any potential contributor be identified for this specific complaint failure mode.Rationale: since no probable root cause was identified for this failure mode, the investigation was not able to identify any corrective actions for this complaint.Based on the reported failure mode, supplier quality notifications were sent to the vendor (b)(4) to notify the vendor of the reported failure mode (anesthesia ineffective).Likewise, the complaint will be entered into the complaint management system and will be tracked & trended for future occurrences and reviewed/investigated through the quality data analysis (qda) process if an adverse trend is identified.
 
Event or Problem Description
It was reported that the anesthesia was ineffective and wore off quickly with a bd tray spn whit27g3.5 l/b-d/e.This occurred on 3 separate occasions, however, the date/time and or patient information is unknown.The following information was provided by the initial reporter: it was reported that the anesthesia was ineffective and wore off quickly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD TRAY SPN WHIT27G3.5 L/B-D/E
Common Device Name
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
CAREFUSION, INC
400 east foster rd
mannford OK 74044
MDR Report Key8865802
Report Number1625685-2019-00085
Device Sequence Number19285594
Product Code CAZ
UDI-Device Identifier00382904056995
UDI-Public00382904056995
Combination Product (Y/N)N
PMA/510(K) Number
DISCRETION
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/30/2020
Device Catalogue Number405699
Device Lot Number0001270921
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/22/2019
Initial Report FDA Received Date08/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-