Reported event: an event regarding fretting involving a metal head which mated with an exeter stem was reported.The event was confirmed through material analysis of the returned product.Method & results: device evaluation and results: visual inspection was performed as part of the material analysis report (mar), dated 10 december 2019.This inspection indicated: the stem and head were returned in the assembled condition.Debris was observed on the head.A material analysis has been performed.The report concluded: the damage present on the articulating surface of the insert was consistent with burnishing, scratching, and third body indentations; which are common damage modes of uhmwpe.The yellow discoloration on the insert was consistent with absorption of synovial fluid.The damage present on the stem was consistent with the cement mantle breakdown process and the collected debris was consistent with an oxide, corrosion product, and biological material.The base material of stem and head were consistent with their drawings.Clinician review: a review of the provided medical records by a clinical consultant stated the following comment: need operative reports, office/clinical reports, serial x-rays, and examination of the explanted components.Device history review: indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: visual inspection was performed as part of the material analysis report (mar), dated 10 december 2019.This inspection indicated: the stem and head were returned in the assembled condition.Debris was observed on the head.A material analysis has been performed.The report concluded: the damage present on the articulating surface of the insert was consistent with burnishing, scratching, and third body indentations; which are common damage modes of uhmwpe.The yellow discoloration on the insert was consistent with absorption of synovial fluid.The damage present on the stem was consistent with the cement mantle breakdown process and the collected debris was consistent with an oxide, corrosion product, and biological material.The exact cause of the event cannot be confirmed as insufficient information was provided.Further information such as operative reports, office/clinical reports, serial x-rays, and examination of the explanted components are required to complete the investigation for determining a root cause.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|