BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO (US) 2.75/26; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 401761
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/09/2019
Type of Reportable Event Malfunction
Event or Problem Description
An orsiro drug-eluting stent system was selected for use.The orsiro could not cross the target lesion in the lad.
 
Additional Manufacturer Narrative
Neither the complaint instrument nor the angiographic material was provided for analysis.Therefore no technical investigation on the subject could be performed.The manufacturing history was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentations confirmed that the instrument was manufactured according to specifications and passed all in process and final inspections.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.The final root cause is most likely related to external factors during procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO (US) 2.75/26
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key8866652
Report Number1028232-2019-03201
Device Sequence Number14276389
Product Code NIQ
UDI-Device Identifier07640130439160
UDI-Public07640130439160
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/03/2020
Device Model Number401761
Device Catalogue NumberSEE MODEL NO.
Device Lot Number09184376
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/26/2019
Supplement Date Received by Manufacturer09/27/2019
Initial Report FDA Received Date08/06/2019
Supplement Report FDA Received Date09/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-