| Model Number |
367019 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/11/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The pantera leo balloon catheter was selected for treatment of a severely calcified lesion in a tortuous left coronary vessel.During inflation the balloon ruptured at 14 atm.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned instrument was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation of the returned device revealed a microscopic leakage and scratches on the balloon surface nearby the damage site.It seems likely that the damage of the balloon surface was caused by a hard, sharpedged object pressing against the balloon from the outside.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test.We can therefore confirm that the instrument was delivered in a leakproof condition.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The root cause is most likely related to the patients anatomy.
|
| |
|